Mouse EPO/Erythropoietin enzyme-linked immunoassay kit
Usd: 569 Special Discount
Specification
Safety datasheet
Overview
Product Name
Mouse EPO/Erythropoietin enzyme-linked immunoassay kit
Product Type
Elisa Kit
Sensitivity
0.07 pg/ml (10 μl)
| pg/ml | O.D. | Average | Corrected | |
|---|---|---|---|---|
| 0.00 | 0.0178 | 0.0135 | 0.0157 | |
| 2.74 | 0.0304 | 0.0264 | 0.0284 | 0.0128 |
| 8.23 | 0.0474 | 0.0548 | 0.0511 | 0.0355 |
| 24.69 | 0.1433 | 0.1391 | 0.1412 | 0.1256 |
| 74.07 | 0.4046 | 0.3610 | 0.3828 | 0.3672 |
| 222.22 | 1.1100 | 1.0450 | 1.0775 | 1.0619 |
| 666.67 | 2.6610 | 2.6220 | 2.6415 | 2.6259 |
| 2000.00 | 4.1867 | 4.2902 | 4.2385 | 4.2228 |
Precision
| Intra-assay Precision | Inter-assay Precision | |||||
| Sample Number | S1 | S2 | S3 | S1 | S2 | S3 |
| 22 | 22 | 22 | 6 | 6 | 6 | |
| Average (pg/ml) | 32.9 | 169.3 | 528.9 | 33.9 | 176.3 | 571.3 |
| Standard Deviation | 0.9 | 3.3 | 20.1 | 0.9 | 9.0 | 27.4 |
| Coefficient of Variation (%) | 2.9 | 1.9 | 3.8 | 2.7 | 5.1 | 4.8 |
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.
Spike Recovery
The spike recovery was evaluated by spiking 3 levels of mouse EPO/Erythropoietin into health mouse serum sample. The un-spiked serum was used as blank in this experiment.
The recovery ranged from 74% to 123% with an overall mean recovery of 102%.
Sample Value
| Sample Matrix | Sample Evaluated | Range (pg/ml) | Detectable (%) | Mean of Detectable (pg/ml) |
|---|---|---|---|---|
| Serum | 30 | 6.43-488.96 | 100 | 71.25 |
Serum/Plasma – Thirty samples from apparently healthy mice were evaluated for the presence of EPO/Erythropoietin in this assay. No medical histories were available for the donors. n.d. = non-detectable. Samples measured below the sensitivity are considered to be non-detectable.
Experiment Type
Species Reactivity
Detection Method
Operation Duration
Standard Curve Gradient
Incubation Count
2
Detectable Sample
Dosing Volume
Product Features
Storage Condition
Store at 2-8 ℃
Target
Function
Same Target and Pathway Products