Mouse GP130/IL16ST enzyme-linked immunoassay kit
Usd: 499 Special Discount
Specification
Safety datasheet
Overview
Product Name
Mouse GP130/IL16ST enzyme-linked immunoassay kit
Product Type
Elisa Kit
Sensitivity
16.07 pg/ml (10 μl)
| pg/ml | O.D. | Average | Corrected | |
|---|---|---|---|---|
| 0.00 | 0.0301 | 0.0346 | 0.0324 | |
| 312.50 | 0.1019 | 0.1033 | 0.1026 | 0.0703 |
| 625.00 | 0.1807 | 0.1666 | 0.1737 | 0.1413 |
| 1250.00 | 0.3318 | 0.3126 | 0.3222 | 0.2899 |
| 2500.00 | 0.6013 | 0.5963 | 0.5988 | 0.5665 |
| 5000.00 | 0.9942 | 1.0350 | 1.0146 | 0.9823 |
| 10000.00 | 1.9440 | 1.9230 | 1.9335 | 1.9012 |
| 20000.00 | 3.4350 | 3.4580 | 3.4465 | 3.4142 |
Precision
| Intra-assay Precision | Inter-assay Precision | |||||
| Sample Number | S1 | S2 | S3 | S1 | S2 | S3 |
| 22 | 22 | 22 | 6 | 6 | 6 | |
| Average (pg/ml) | 510.1 | 2405.5 | 7549.3 | 413.6 | 2054.1 | 5914.5 |
| Standard Deviation | 37.4 | 117.9 | 524.8 | 5.7 | 89.6 | 198.6 |
| Coefficient of Variation (%) | 7.3 | 4.9 | 7.0 | 2.5 | 2.2 | 2.9 |
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty-two times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.
Spike Recovery
The spike recovery was evaluated by spiking 3 levels of mouse GP130/IL16ST into health mouse serum sample. The un-spiked serum was used as blank in this experiment.
The recovery ranged from 80% to 100% with an overall mean recovery of 97%.
Sample Value
| Sample Matrix | Sample Evaluated | Range (ng/ml) | Detectable (%) | Mean of Detectable (ng/ml) |
|---|---|---|---|---|
| Serum | 30 | 10.56-16.10 | 100 | 13.46 |
Serum/Plasma– Thirty samples from apparently healthy mice were evaluated for the presence of mouse GP130/IL16ST in this assay. No medical histories were available for the donors. n.d. = non-detectable. Samples measured below the sensitivity are considered to be non-detectable.
Experiment Type
Species Reactivity
Detection Method
Operation Duration
Standard Curve Gradient
Incubation Count
2
Detectable Sample
Dosing Volume
Product Features
Storage Condition
Store at 2-8 ℃
Target
UNIPROT
Function
Same Target and Pathway Products