Mouse/Rat IGFBP6 enzyme-linked immunoassay kit(one step)
Usd: 619 Special Discount
Specification
Safety datasheet
Overview
Product Name
Mouse/Rat IGFBP6 enzyme-linked immunoassay kit(one step)
Product Type
Elisa Kit
Sensitivity
0.29ng/ml (10 μl)
| ng/ml | O.D. | Average | Corrected | |
|---|---|---|---|---|
| 0.00 | 0.0064 | 0.0073 | 0.0069 | |
| 4.69 | 0.0283 | 0.0271 | 0.0277 | 0.0209 |
| 9.38 | 0.0560 | 0.0568 | 0.0564 | 0.0496 |
| 18.75 | 0.1186 | 0.1181 | 0.1184 | 0.1115 |
| 37.50 | 0.2680 | 0.2830 | 0.2755 | 0.2687 |
| 75.00 | 0.6413 | 0.6167 | 0.6290 | 0.6222 |
| 150.00 | 1.4230 | 1.4820 | 1.4525 | 1.4457 |
| 300.00 | 2.9660 | 2.9030 | 2.9345 | 2.9277 |
Precision
| Intra-assay Precision | Inter-assay Precision | |||||
| Sample Number | S1 | S2 | S3 | S1 | S2 | S3 |
| 22 | 22 | 22 | 6 | 6 | 6 | |
| Average (ng/ml) | 5.7 | 26.9 | 90.6 | 6.0 | 26.8 | 84.1 |
| Standard Deviation | 0.3 | 1.5 | 7.1 | 0.3 | 1.1 | 4.1 |
| Coefficient of Variation (%) | 3.2 | 3.4 | 4.7 | 4.9 | 4.1 | 4.9 |
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty-two times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.
Spike Recovery
The spike recovery was evaluated by spiking 3 levels of mouse/rat IGFBP6 into health mouse/rat serum sample. The un-spiked serum was used as blank in this experiment.
The recovery ranged from 83% to 110% with an overall mean recovery of 93%.
Sample Value
| Sample Matrix | Sample Evaluated | Range (ng/ml) | Detectable (%) | Mean of Detectable (ng/ml) |
|---|---|---|---|---|
| Mouse Serum | 30 | 16.56-204.61 | 100 | 103.01 |
| Rat Serum | 30 | 129.40-316.34 | 100 | 229.71 |
Serum/Plasma – Thirty samples from apparently healthy mice and thirty samples from apparently healthy rats were evaluated for the presence of mouse/rat IGFBP6 in this assay. No medical histories were available for the donors.n.d. = non-detectable. Samples measured below the sensitivity are considered to be non-detectable.
Experiment Type
Species Reactivity
Detection Method
Operation Duration
Standard Curve Gradient
Incubation Count
1
Detectable Sample
Dosing Volume
Product Features
Storage Condition
Store at 2-8 ℃
Target
Alias
IBP 6
IBP-6
IBP6
IGF binding protein 6
IGF-binding protein 6
IGFBP 6
IGFBP-6
IGFBP6
Insulin like growth factor binding protein 6
Insulin-like growth factor-binding protein 6
Function
Products with the Same Target and Pathway