Human CD66e/CEACAM-5 enzyme-linked immunoassay kit
Usd: 499 Special Discount
Specification
Safety datasheet
Overview
Product Name
Human CD66e/CEACAM-5 enzyme-linked immunoassay kit
Product Type
Elisa Kit
Sensitivity
2.60 pg/ml (50 ul)
| pg/ml | O.D. | Average | Corrected | |
|---|---|---|---|---|
| 0.00 | 0.0397 | 0.0395 | 0.0396 | |
| 15.63 | 0.1158 | 0.1122 | 0.1140 | 0.0744 |
| 31.25 | 0.1874 | 0.1708 | 0.1791 | 0.1395 |
| 62.50 | 0.3159 | 0.3148 | 0.3154 | 0.2758 |
| 125.00 | 0.5713 | 0.5489 | 0.5601 | 0.5205 |
| 250.00 | 1.0140 | 0.9802 | 0.9971 | 0.9575 |
| 500.00 | 1.6900 | 1.7070 | 1.6985 | 1.6589 |
| 1000.00 | 2.8040 | 2.7650 | 2.7845 | 2.7449 |
Precision
| Intra-assay Precision | Inter-assay Precision | |||||
| Sample Number | S1 | S2 | S3 | S1 | S2 | S3 |
| 22 | 22 | 22 | 6 | 6 | 6 | |
| Average (pg/ml) | 19.1 | 99.5 | 302.4 | 22.6 | 98.6 | 275.8 |
| Standard Deviation | 1.2 | 3.8 | 12.9 | 1.0 | 3.9 | 11.3 |
| Coefficient of Variation (%) | 6.2 | 3.8 | 4.3 | 4.4 | 3.9 | 4.1 |
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty-two times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.
Spike Recovery
The spike recovery was evaluated by spiking 3 levels of human CD66e/CEACAM5 into health human serum sample. The un-spiked serum was used as blank in this experiment.
The recovery ranged from 99% to 115% with an overall mean recovery of 105%.
Sample Value
| Sample Matrix | Sample Evaluated | Range (pg/ml) | Detectable (%) | Mean of Detectable (pg/ml) |
|---|---|---|---|---|
| serum | 30 | 35.50-300.61 | 100.0 | 112.69 |
Serum/Plasma – Thirty samples from apparently healthy volunteers were evaluated for the presence of human CD66e/CEACAM5 in this assay. No medical histories were available for the donors. n.d. = non-detectable. Samples measured below the sensitivity are considered to be non-detectable.
Experiment Type
Species Reactivity
Detection Method
Operation Duration
Standard Curve Gradient
Incubation Count
2
Detectable Sample
Dosing Volume
Product Features
Storage Condition
Store at 2-8 ℃
Target