Human Galectin3 enzyme-linked immunoassay kit
Usd: 499 Special Discount
Specification
Safety datasheet
Overview
Product Name
Human Galectin3 enzyme-linked immunoassay kit
Product Type
Elisa Kit
Sensitivity
125.02 pg/ml (50 μl);404.56 pg/ml (10 μl);
| pg/ml | O.D. | Average | Corrected | |
|---|---|---|---|---|
| 0.00 | 0.0186 | 0.0177 | 0.0182 | |
| 312.50 | 0.0266 | 0.0256 | 0.0261 | 0.0080 |
| 625.00 | 0.0389 | 0.0386 | 0.0388 | 0.0206 |
| 1250.00 | 0.0849 | 0.0847 | 0.0848 | 0.0667 |
| 2500.00 | 0.2597 | 0.2528 | 0.2563 | 0.2381 |
| 5000.00 | 0.7995 | 0.7769 | 0.7882 | 0.7701 |
| 10000.00 | 2.1670 | 2.1290 | 2.1480 | 2.1299 |
| 20000.00 | 3.8230 | 3.8410 | 3.8320 | 3.8139 |
Precision
| Intra-assay Precision | Inter-assay Precision | |||||
| Sample Number | S1 | S2 | S3 | S1 | S2 | S3 |
| 22 | 22 | 22 | 6 | 6 | 6 | |
| Average (pg/ml) | 341.7 | 1933.2 | 5656.0 | 396.7 | 1924.8 | 5724.1 |
| Standard Deviation | 25.4 | 50.6 | 153.1 | 30.8 | 36.8 | 93.0 |
| Coefficient of Variation (%) | 7.4 | 2.6 | 2.7 | 7.8 | 1.9 | 1.6 |
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty-two times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.
Spike Recovery
The spike recovery was evaluated by spiking 3 levels of human Galectin3 into health human serum sample. The un-spiked serum was used as blank in this experiment.
The recovery ranged from 90% to 112% with an overall mean recovery of 108%.
Sample Value
| Sample Matrix | Sample Evaluated | Range (ng/ml) | Detectable (%) | Mean of Detectable (ng/ml) |
|---|---|---|---|---|
| Serum | 30 | 60.49-135.27 | 100.0 | 93.91 |
Serum/Plasma– Thirty samples from apparently healthy volunteers were evaluated for the presence of human Galectin3 in this assay. No medical histories were available for the donors. n.d. = non-detectable. Samples measured below the sensitivity are considered to be non-detectable.
Experiment Type
Species Reactivity
Detection Method
Operation Duration
Standard Curve Gradient
Incubation Count
2
Detectable Sample
Dosing Volume
Product Features
Storage Condition
Store at 2-8 ℃
Target
UNIPROT
Function