Human LAG-3 enzyme-linked immunoassay kit
Usd: 619 Special Discount
Specification
Safety datasheet
Overview
Product Name
Human LAG-3 enzyme-linked immunoassay kit
Product Type
Elisa Kit
Sensitivity
3.40 pg/ml (50 μl) ;36.81 pg/ml (10 μl)
| pg/ml | O.D. | Average | Corrected | |
|---|---|---|---|---|
| 0.00 | 0.0152 | 0.0139 | 0.0139 | |
| 156.25 | 0.0949 | 0.0900 | 0.0925 | 0.0786 |
| 312.50 | 0.1725 | 0.1668 | 0.1697 | 0.1558 |
| 625.00 | 0.3285 | 0.3315 | 0.3300 | 0.3161 |
| 1250.00 | 0.6285 | 0.6328 | 0.6307 | 0.6168 |
| 2500.00 | 1.1310 | 1.1440 | 1.1375 | 1.1236 |
| 5000.00 | 1.9550 | 2.0120 | 1.9835 | 1.9696 |
| 10000.00 | 3.0790 | 3.0990 | 3.0890 | 3.0751 |
Precision
| Intra-assay Precision | Inter-assay Precision | |||||
| Sample Number | S1 | S2 | S3 | S1 | S2 | S3 |
| 22 | 22 | 22 | 6 | 6 | 6 | |
| Average (pg/ml) | 222.9 | 1049.2 | 3198.7 | 190.1 | 934.2 | 2834.4 |
| Standard Deviation | 8.9 | 31.4 | 153.4 | 7.3 | 20.1 | 92.9 |
| Coefficient of Variation (%) | 4.0 | 3.0 | 4.8 | 3.9 | 2.1 | 3.3 |
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty-two times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.
Spike Recovery
The spike recovery was evaluated by spiking 3 levels of human LAG-3 into health human serum sample. The un-spiked serum was used as blank in this experiment.
The recovery ranged from 81% to 118% with an overall mean recovery of 95%.
Sample Value
| Sample Matrix | Sample Evaluated | Range (pg/ml) | Detectable (%) | Mean of Detectable (pg/ml) |
|---|---|---|---|---|
| Serum | 30 | 474.27-6835.60 | 100 | 2298.95 |
Serum/Plasma – Thirty samples from apparently healthy volunteers were evaluated for the presence of human LAG-3 in this assay. No medical histories were available for the donors. n.d. = non-detectable. Samples measured below the sensitivity are considered to be non-detectable.
Experiment Type
Species Reactivity
Detection Method
Operation Duration
Standard Curve Gradient
Incubation Count
2
Detectable Sample
Dosing Volume
Product Features
Storage Condition
Store at 2-8 ℃
Target
UNIPROT
Function