Human CD30 Ligand/TNFSF8/CD 30-L enzyme-linked immunoassay kit
Usd: 499 Special Discount
Specification
Safety datasheet
Overview
Product Name
Human CD30 Ligand/TNFSF8/CD 30-L enzyme-linked immunoassay kit
Product Type
Elisa Kit
Sensitivity
2.05 pg/ml (50 μl);9.27 pg/ml (10 μl);
| pg/ml | O.D. | Average | Corrected | |
|---|---|---|---|---|
| 0.00 | 0.0596 | 0.0705 | 0.0651 | |
| 31.25 | 0.1439 | 0.1665 | 0.1552 | 0.0902 |
| 62.50 | 0.2245 | 0.2592 | 0.2419 | 0.1768 |
| 125.00 | 0.3960 | 0.4307 | 0.4134 | 0.3483 |
| 250.00 | 0.7496 | 0.8006 | 0.7751 | 0.7101 |
| 500.00 | 1.3630 | 1.3900 | 1.3765 | 1.3115 |
| 1000.00 | 2.3260 | 2.5400 | 2.4330 | 2.3680 |
| 2000.00 | 3.9380 | 4.0608 | 3.9994 | 3.9344 |
Precision
| Intra-assay Precision | Inter-assay Precision | |||||
| Sample Number | S1 | S2 | S3 | S1 | S2 | S3 |
| 22 | 22 | 22 | 6 | 6 | 6 | |
| Average (pg/ml) | 24.8 | 166.3 | 576.3 | 28.0 | 174.4 | 581.4 |
| Standard Deviation | 1.9 | 8.2 | 24.8 | 1.5 | 3.7 | 14.1 |
| Coefficient of Variation (%) | 7.7 | 4.9 | 4.3 | 5.5 | 2.1 | 2.4 |
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty-two times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.
Spike Recovery
The spike recovery was evaluated by spiking 3 levels of human CD30 Ligand/TNFSF8/CD 30-L into health human serum sample. The un-spiked serum was used as blank in this experiment.
The recovery ranged from 89% to 110% with an overall mean recovery of 101%.
Sample Value
| Sample Matrix | Sample Evaluated | Range (pg/ml) | Detectable (%) | Mean of Detectable (pg/ml) |
|---|---|---|---|---|
| Human Serum | 30 | 197.54-394.71 | 100 | 297.34 |
Serum/Plasma– Thirty samples from apparently healthy volunteers/mouse serum were evaluated for the presence of human CD30 Ligand/TNFSF8/CD 30-L in this assay. No medical histories were available for the donors. n.d. = non-detectable. Samples measured below the sensitivity are considered to be non-detectable.
Experiment Type
Species Reactivity
Detection Method
Operation Duration
Standard Curve Gradient
Incubation Count
2
Detectable Sample
Dosing Volume
Product Features
Storage Condition
Store at 2-8 ℃
Target
UNIPROT
Function