Human CD40 enzyme-linked immunoassay kit
Usd: 499 Special Discount
Specification
Safety datasheet
Overview
Product Name
Human CD40 enzyme-linked immunoassay kit
Product Type
Elisa Kit
Sensitivity
0.34 pg/ml (50 μl)
| pg/ml | O.D. | Average | Corrected | |
|---|---|---|---|---|
| 0.00 | 0.0222 | 0.0225 | 0.0224 | |
| 1.37 | 0.0257 | 0.0279 | 0.0268 | 0.0045 |
| 4.12 | 0.0406 | 0.0415 | 0.0411 | 0.0187 |
| 12.35 | 0.0851 | 0.0860 | 0.0856 | 0.0633 |
| 37.04 | 0.2245 | 0.2166 | 0.2206 | 0.1982 |
| 111.11 | 0.6181 | 0.6157 | 0.6169 | 0.5946 |
| 333.33 | 1.7090 | 1.7460 | 1.7275 | 1.7052 |
| 1000.00 | 3.8810 | 3.9440 | 3.9125 | 3.8902 |
Precision
| Intra-assay Precision | Inter-assay Precision | |||||
| Sample Number | S1 | S2 | S3 | S1 | S2 | S3 |
| 22 | 22 | 22 | 6 | 6 | 6 | |
| Average (pg/ml) | 23.6 | 121.8 | 373.3 | 23.7 | 115.7 | 342.3 |
| Standard Deviation | 0.5 | 2.8 | 8.2 | 1.0 | 3.0 | 12.5 |
| Coefficient of Variation (%) | 2.1 | 2.3 | 2.2 | 4.1 | 2.6 | 3.7 |
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty-two times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.
Spike Recovery
The spike recovery was evaluated by spiking 3 levels of human CD40 into health human serum sample. The un-spiked serum was used as blank in this experiment.
The recovery ranged from 91% to 121% with an overall mean recovery of 110%.
Sample Value
| Sample Matrix | Sample Evaluated | Range (pg/ml) | Detectable (%) | Mean of Detectable (pg/ml) |
|---|---|---|---|---|
| Serum | 30 | 208.10-361.52 | 100 | 281.37 |
Serum/Plasma – Thirty samples from apparently healthy volunteers were evaluated for the presence of human CD40 in this assay. No medical histories were available for the donors. n.d. = non-detectable. Samples measured below the sensitivity are considered to be non-detectable.
Experiment Type
Species Reactivity
Detection Method
Operation Duration
Standard Curve Gradient
Incubation Count
2
Detectable Sample
Dosing Volume
Product Features
Storage Condition
Store at 2-8 ℃
Target
UNIPROT
Function
Same Target and Pathway Products