Human Non-Phospho-Tau(T217) enzyme-linked immunoassay kit
Usd: 619 Special Discount
Specification
Safety datasheet
Overview
Product Name
Human Non-Phospho-Tau(T217) enzyme-linked immunoassay kit
Product Type
Elisa Kit
Sensitivity
18.18 pg/ml (50 ul)
| 0.00 | 0.0831 | 0.0802 | 0.0817 | |
|---|---|---|---|---|
| 78.13 | 0.1625 | 0.1516 | 0.1571 | |
| 0.0754 | 156.25 | 0.2404 | 0.2264 | 0.2334 |
| 0.1518 | 312.50 | 0.3893 | 0.3644 | 0.3769 |
| 0.2952 | 625.00 | 0.6965 | 0.6539 | 0.6752 |
| 0.5936 | 1250.00 | 1.1970 | 1.2180 | 1.2075 |
| 1.1259 | 2500.00 | 2.1600 | 2.1580 | 2.1590 |
| 2.0774 | 5000.00 | 3.8800 | 3.7210 | 3.8005 |
| 3.7189 | ||||
Precision
| Intra-assay Precision | Inter-assay Precision | |||||
| Sample Number | S1 | S2 | S3 | S1 | S2 | S3 |
| 22 | 22 | 22 | 6 | 6 | 6 | |
| Average (pg/ml) | 82.6 | 420.0 | 1338.6 | 96.5 | 427.8 | 1410.5 |
| Standard Deviation | 6.4 | 22.1 | 98.4 | 4.4 | 17.2 | 98.0 |
| Coefficient of Variation (%) | 6.5 | 5.3 | 7.4 | 4.5 | 4.0 | 7.0 |
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty-two times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.
Spike Recovery
The spike recovery was evaluated by spiking 3 levels of human Non-Phospho-Tau(T217) into health human serum sample. The un-spiked serum was used as blank in this experiment.
The recovery ranged from 82% to 115% with an overall mean recovery of 97%.
Sample Value
| Sample Matrix | Sample Evaluated | Range (pg/ml) | Detectable (%) | Mean of Detectable (pg/ml) |
|---|---|---|---|---|
| Serum | 30 | n.d.-1363.44 | 80 | 489.53 |
Serum/Plasma – Thirty samples from apparently healthy volunteers were evaluated for the presence of human Non-Phospho-Tau(T217) in this assay. No medical histories were available for the donors. n.d. = non-detectable. Samples measured below the sensitivity are considered to be non-detectable.
Experiment Type
Species Reactivity
Detection Method
Operation Duration
Standard Curve Gradient
Incubation Count
2
Detectable Sample
Dosing Volume
Product Features
Storage Condition
Store at 2-8 ℃
Target
UNIPROT
Same Target and Pathway Products